CentraLink Data Management System V14 (Analyser software for in vitro diagnostic medical device (IVD))Siemens Material Numbers: 10811600, 10811601, 10814296, 10814298, 10814877, 10814880, 10814881, 10811741, 10811742, 10811743, 10815474, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10815474 & 10811596ARTG Number: 224615 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00926-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-08-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has identified a task scheduler issue with the centralink data management system, software version 14. the issue may cause the software to stop executing commands, including uploading validated results to the lis. the issue is related to an internal software timer that overflows after 24 days. automated rules triggered by task schedulers or by actions in the centralink user interface may also be affected. the specific consequences of discontinued command execution will vary:1) automated rules will stop running. results may be released to the lis that should have been held based on the automated rules.2) tasks to upload results and download orders to/from the lis will stop running causing a delay in test processing. 3) data in memory may not be updated causing data displayed on reports to not be up to date. however, the lis information would be correct.4) a database backup may not occur when scheduled.
  • 조치
    Siemens Healthcare Diagnostics is advising customers to log out of user sessions and restart the CentraLink services and translators as soon as it is convenient. Then periodically (more often than every 24 days) using the procedure provided to restart the CentraLink Data Management System. Siemens is currently investigating further corrective actions. This action has been closed-out on 18/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    CentraLink Data Management System V14 (Analyser software for in vitro diagnostic medical device (IVD))Siemens Material Numbers: 10811600, 10811601, 10814296, 10814298, 10814877, 10814880, 10814881, 10811741, 10811742, 10811743, 10815474, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10815474 & 10811596ARTG Number: 224615
  • Manufacturer

Manufacturer