Centricity PACS RA1000 Workstation versions 3.2 and higher (Picture archiving and communication system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare IITS Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00442-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified that in a reject image workflow where multiple studies are simultaneously opened on the centricity pacs ra1000 workstation and images are rejected from the active displayed study, images from the previously displayed study may also be rejected. this issue can occur with a previously displayed study that remains open in the background.There is a delrejimg task in centricity pacs versions 3.2 and higher, that if enabled and running, in conjunction with the inadvertently rejected image issue, poses a potential patient safety hazard due to data loss. if the delrejimg task runs, the inadvertently rejected images will be permanently deleted from pacs. if you do not have the delrejimg task running, there is no potential hazard since the inadvertently rejected images remain on the system. the deletion of the inadvertently rejected images could result in missed intervention or mis-diagnosis. no patient injuries have been reported.
  • 조치
    A future release of Centricity PACS will address the unintentional rejected image issue. Until the permanent fix for this issue is deployed, GE recommends customers implement workarounds identified in the customer letter. Also, a GE Healthcare service representative will remove the DelRejImg task from the task table in the Centricity PACS database. This will remove the rejected image deletion feature, and eliminate the hazard situation. Thia action has been closed-out on 28/01/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Centricity PACS RA1000 Workstation versions 3.2 and higher (Picture archiving and communication system)Software versions: 3.2, 3.2.0.1, 3.2.0.2, 3.2.1, 3.2.2, 3.2.2.1, 3.2.2.2, 3.2.2.3, 3.2.6, 3.2 SP7, 3.2 SP8, 4.0, 4.0.1 & 4.0 SP3ARTG number: 162031
  • Manufacturer

Manufacturer