Certegra Workstation used in conjunction with Medrad Stellant CT Injection System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Imaxeon Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01064-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medrad has identified that an internal cable connection in the all-in-one computer component of certain certegra workstations used in conjunction with the medrad stellant ct injection system may potentially exhibit thermal events such as smoke, sparks or fire. this is a result of degradation over time of the shielding within the connection. if such an event occurs, a delay of procedure, coughing, nausea, difficulty breathing, or first or second degree burns may occur. there have been no patient or user injuries reported as a result of this situation.
  • 조치
    Imaxeon will be arranging for impacted Workstations to be replaced. In the interim, users are advised to take the following actions. 1. While computers are normally warm during use, remain vigilant in checking the back of the Workstation for signs of smoke or other thermal activity while the system is in use. If smoke or other thermal activity is observed, unplug the power cord and discontinue use of your Workstation until it is replaced. 2. Power down the system when you are closing your CT suite for the day or when the CT suite will not be staffed for an extended time and then power it back up prior to your next contrast enhanced CT scan. Interim Instructions for Use with further details will be provided with the customer letter. This action has been closed-out on 22/02/2017.