Cobas IT 3000 Haematology Performance Suite, keyboard differential keyboard (with Cobas IT 3000 versions 2.03.08 to 2.04.00). An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00782-3
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-08-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is a potential that the analyte comments may be deleted if the specific analyte is not counted in the mask. this risk is avoided in the case that the user makes comments against the “blood film comments” analyte for reporting blood film comments. the differential keyboard orders a profile that contains all of the additional blood cell types that can be manually counted. in many instances these additional cell types will not be identified in the count and will not receive a result (eg blasts, myelocytes, etc). the expected behaviour is that if a cell is not counted, then these analytes will be removed from the profile when the final count is accepted.
  • 조치
    Roche is providing work around instructions until a software upgrade can be implemented to permanently fix the issue.