cobas omni MGP Reagent when used with cobas MPX assay. An in vitro diagnostic medical device (IVD)(for use on the cobas 6800/8800 systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00744-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-06-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Roche diagnostics has observed an increased rate of invalid internal control (ic) results for cobas mpx when used with certain lots of the cobas omni mgp reagent. this issue affects only the blood screening assays, cobas mpx. for the non-blood screening assays, these lots of cobas omni mgp reagent continue to meet specifications, and no performance issues have been identified. the cobas 6800/8800 systems will correctly flag a sample result as invalid if the internal control is invalid. there is no risk of generating incorrect patient results as a consequence of this issue. all previously generated valid results and runs remain valid.
  • 조치
    Roche is advising customers to restrict the use of affected cobas omni MGP Reagent lots to non-blood screening assays only (CMV; CT/NG; HPV), and discontinue the use of these specific lots with the cobas MPX.

Device

  • 모델명 / 제조번호(시리얼번호)
    cobas omni MGP Reagent when used with cobas MPX assay. An in vitro diagnostic medical device (IVD)(for use on the cobas 6800/8800 systems)cobas omni MGP Reagent Material Number: 06997546190Pack Size: 480 testsLot Numbers: W17311, X05638, X05639, X16159ARTG Number: 175163
  • Manufacturer

Manufacturer