Colleague Volumetric Infusion Pumps 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Baxter Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00384-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-03-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During 2011, it was identified that a quantity of pumps fitted with two ceramic 1.6a slow-blow fuses in the rear housing was failing prematurely. investigation led to a change of the fuse type fitted during future manufacture and service, from a ceramic fuse to a glass fuse, which has a different rupture characteristic. the failure rate of this reported issue has subsequently declined following the replacement with a glass fuse in newly manufactured pumps. if the fuses fail the device cannot charge the batteries even when connected to a.C. power. failure of the fuses while the pump is in operation would result in the pump switching from a.C. power to battery power and the a.C. icon would extinguish. the pump would continue to run on battery power with corresponding battery discharge alarms.
  • 조치
    Customers are asked to examine the two fuses installed in the rear of the pump which can be accessed by slotted fuse covers to establish if the fuses are glass or ceramic type. Ceramic fuses should be replaced with the glass fuses which will be supplied by Baxter. For Baxter owned pumps placed with customers, Baxter will carry out this activity at the next service.

Device

  • 모델명 / 제조번호(시리얼번호)
    Colleague Volumetric Infusion Pumps Product Codes: 2M91617 & 2M91637 ARTG Number: 177494