Communication Engine in Plum A+ and A+3 Infusion Pump 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hospira Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00140-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2014-02-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • 데이터 추가 비고
  • 원인
    Some plum a+/a+3 devices contain a communication engine which communicates wirelessly with hospira mednet over a facility's local area network (lan). hospira has received several reports indicating that the communication engine in some plum a+/a+3 devices experiences difficulty downloading a custom drug library as indicated by an appropriate error message.
  • 조치
    Hospira are advising users that if a custom drug library fails to download and experiences an error message, please report the issue to Hospira. Hospira will make arrangements to identify and correct the Communication Engine.

Device

  • 모델명 / 제조번호(시리얼번호)
    Communication Engine in Plum A+ and A+3 Infusion Pump Plum A+Pump: 12391 & 20792Communication Engine Modules:20677 & 20791 (respectively)Plum A+3Pump: 20678Communication Engine Modules: 20678ARTG Number: 138109
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source