Compact Airway Modules E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV & E-CAiOVX (used with CARESCAPE Monitors B850 and B650, Avance and Aisys Anesthesia Carestations, Engstrom Respiratory Carestation, or S/5 Anesthesia and Critical Care Monitors) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00488-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has recently become aware of a potential safety issue due to gas sampling pump failure associated with the compact airway gas modules. loosening of the gas sampling pump hex screw in the module pump may lead to pump failure that may cause loss of respiratory airway gas parameter monitoring. if this issue occurs, the monitor will show an alarm note on monitor screen of either "low gas sample flow" or "check sample gas out".
  • 조치
    GE Healthcare will implement a correction for the affected modules. If the monitor is showing an alarm of “Low gas sample flow” or “Check sample gas out”, users are advised to check the following before continued use: 1. Check the water trap is not full or occluded. If full or occluded, empty/change the water trap. 2. Check gas sampling adapter at patient airway is not occluded. 3. Check sample line is not occluded. If occluded, change new gas sampling line. 4. Check gas exhaust outlet is not occluded. Ensure there is no blockage preventing the sample gas out flow to scavenging. If the above actions do not remove the alarm message(s), it may be due to the issue described above. Users are advised to remove the equipment from use and contact technical support or local service representative. This action has been closed-out on 02/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Compact Airway Modules E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV & E-CAiOVX (used with CARESCAPE Monitors B850 and B650, Avance and Aisys Anesthesia Carestations, Engstrom Respiratory Carestation, or S/5 Anesthesia and Critical Care Monitors)Affected serial number range # 6704120 - 6778756ARTG Number: 166230