Conmed GS2000 - 50L Abdominal Insufflator(provides CO2 gas distension of surgical cavities via an endoscope, for diagnostic and/or operative endoscopy) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 ConMed Linvatec Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01601-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-12-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The reason for this recall is that the device could cause a patient overpressure situation without any visual or auditory indication or warning from the device. if specific conditions occur, the affected unit could create an overpressure which is transient and reversible, requiring intervention from medical staff. if the device were to fail, the surgical attendants, anaesthetist or doctor would see an over -distension of the pneumoperitoneum or difficulty in patient breathing. the manufacturer has received one adverse event report overseas where the device caused an overpressure situation.
  • 조치
    Conmed is advising users to immediately check their facility for the affected unit(s) and to contact Conmed, who will in turn organise collection of the affected unit and return of the corrected unit. If necessary, Conmed will provide a loaner device whilst the product correction is undertaken.

Device

  • 모델명 / 제조번호(시리얼번호)
    Conmed GS2000 - 50L Abdominal Insufflator(provides CO2 gas distension of surgical cavities via an endoscope, for diagnostic and/or operative endoscopy)Affected Serial Numbers: 90869KZD and 91114CADARTG Number: 233786
  • 의료기기 분류등급
  • Manufacturer

Manufacturer