Contact Detach Infusion Sets ( used in conjunction with a pump to deliver a continuous dose of insulin) Manufactured prior to May 2014 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Australasian Medical & Scientific Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00703-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-07-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has identified a slight increase in reported cases of the steel needle breaking during use for infusions sets manufactured prior to may 2014. in a small sub set of these reported cases, this needle breakage has led to hospitalisation for the management of glucose levels and/or treatment for removal of the needle. needle breakage can result in the interruption of insulin delivery and can cause hyperglycaemia, which if left untreated, can result in diabetic ketoacidosis (dka). in addition to the interruption of insulin, a needle remaining in the body might, in rare cases, require surgical removal if the needle does not itself migrate to the surface where it can be removed with a pair of tweezers.
  • 조치
    In the event consumers wish to carry on with using the old needle they are reminded to follow the current IFU along with specific additional instructions in the customer letter; OR if they prefer users (including members of the public) can phone AMSL for replacement with unaffected stock. This action has been closed-out on 15/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Contact Detach Infusion Sets ( used in conjunction with a pump to deliver a continuous dose of insulin)Manufactured prior to May 2014Multiple Lot Numbers – see attached listARTG Number: 116234
  • Manufacturer

Manufacturer