Conveyor unit (CV-50) of Sysmex XN-9000 automated haematology analysers. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00501-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2018-05-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Roche have been notified by the manufacturer, sysmex corporation, of a potential failure of power up due to a short circuit of the pcb of cv-50 power supply unit in the event a leakage occurs from the instrument or operator error. sysmex have identified that in rare cases, some fluid (approx 50ml) may leak from the xn-10 / xn-20 automated haematology analysers into the conveyor unit itself, resulting in short circuit of the power supply and subsequent malfunction of the conveyor. if this issue occurs, the system will stop and patient samples will require processing by manual measurement.Sysmex has received one complaint worldwide for this issue; none from australian customers to date.
  • 조치
    Roche is advising they will be in contact with customers to arrange for a technician to ensure the gaps on the outer covers of the CV-50 will be sealed with waterproof materials. This will prevent fluid from entering the power supply board.

Device

  • 모델명 / 제조번호(시리얼번호)
    Conveyor unit (CV-50) of Sysmex XN-9000 automated haematology analysers. An in vitro diagnostic medical device (IVD)Material Number: 06510027001ARTG Number: 173886
  • Manufacturer

Manufacturer