Corpatch Easy Paediatric (defibrillator electrodes) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00388-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-05-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An internal investigation showed that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. this malfunction can occur with defibrillation electrodes that are older than 2 years. this could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all. the precise cause for this is still under investigation. the electrodes are being in used in connection with the defibrillators corpuls 08/16, corpuls1 and corpuls3. other defibrillation electrodes manufactured for the corpuls defibrillators are not affected by the problem.
  • 조치
    Device Technologies Australia (DTA) is requesting their customers to isolate the affected electrodes. DTA is contacting the customers with a Good Return Authorisation (GRA) number and arrange collection. Replacement paediatric leads are available and will be provided as soon as possible. In the interim, if no paediatric defibrillation electrodes are available, DTS advises that standard adult electrodes may be used for defibrillation of patients <25kg. This action has been closed-out on 08/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Corpatch Easy Paediatric (defibrillator electrodes)Product Code CP-5120.2 Lot Numbers: 30708-0777, 30822-0778, 31009-0773, 31121-0774, 40107-0771, 40213-0774, 40222-0773, 40320-0770, 40321-0770, 40324-0773, 40522-0779, 40704-0776, 40808-0773, 40923-0776, 41021-0778.ARTG number: 200520