Cortex Screw Ø 3.5 mm, self-tapping, length 28 mm, Stainless Steel 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Synthes Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00948-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-08-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Due to a manufacturing error, a cortex screw ø 3.5 mm, self-tapping, length 30 mm, stainless steel (part number 204.828) with an unfinished thread and malformed tip was packaged and labeled as a cortex screw ø 3.5 mm, length 28 mm, stainless steel (part number 204.028). if the presence of a screw that is 30mm instead of 28mm, with a tip that is malformed, is detected, a marginal surgical delay may arise as an alternative screw is selected. the potential for a delay increases if multiple screws of the incorrect length/malformation are present in the graphic case. if the presence is not detected, the surgeon may attempt to utilize the screw during a surgical procedure. when the surgeon attempts to insert the screw, the diameter of the non-threaded ball-like tip of the screw will not fit into the hole that is drilled. therefore insertion will not be possible.
  • 조치
    Customers are asked to check their inventory immediately for the affected units and if they have any of the affected devices to immediately isolate them. This action has been closed-out on 03/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Cortex Screw Ø 3.5 mm, self-tapping, length 28 mm, Stainless Steel Part Number: 204.828 Lot Numbers: 8573875; 8594066 ARTG Number: 157068