Cortisol Saliva ELISA. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Abacus ALS Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00109-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-02-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, ibl has detected an issue that indicates a degradation in performance of the kit resulting in elevated cortisol values. when testing patient samples with one of the affected lots, the controls and calibrator values will meet specifications; however, due to a degradation in the starting concentration of the calibration standards, the calculated results for individual patient samples may be higher compared to lc-ms/ms measurements and may trigger unnecessary additional testing.A review of additional data comparing the respective kit standards of the lots to lc-ms/ms data it is evident that the lots eco142 to eco143 generate elevated results:- eco142 (passing bablok fit (y=0.009988 + 1.549x)- eco143 (passing bablok fit (y=0.003142 + 1.623x)there has not been any reports of injuries associated with this issue.
  • 조치
    Abacus is advising users to inspect stock and quarantine any remaining stock of the affected batches. Affected product will be replaced with unaffected stock. A review of previously generated patient results is at the discretion of the Laboratory Director.

Device

  • 모델명 / 제조번호(시리얼번호)
    Cortisol Saliva ELISA. An in vitro diagnostic medical device (IVD)Catalogue Number: RE52611Lot Numbers: ECO142, ECO143Expiry Dates: ECO142: 2018-02-28; ECO143: 2018-12-31ARTG Number: 210636 (Abacus ALS - Clinical chemistry hormone IVDs)
  • Manufacturer

Manufacturer