Cranial Screw PlusDrive ø 1.6 mm, Self-Drilling, L 3mm (part of the Low Profile Neuro System intended for use in selective trauma of the midface and craniofacial skeleton) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00256-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The affected part number and lot numbers are out of specification at the thread. this is related to a non-conformance where the thread height of these screws is under-sized. it was also identified that the cross section of the affected area is less than that of conforming screws such that the thread height of the screws is under-sized.The decreased thread height may alter the self-drilling feature and result in difficulty inserting the screw. if the surgeon finds that the screw is not inserting as anticipated, the surgeon may choose to use a different screw or drill a pilot hole to insert the screw. attempting to insert additional screws or possibly drill with another instrument could result in a surgical delay. should the user be able to insert the screw the decreased thread height may significantly reduce the retention ability of the screw in the bone. these issues may require medical or surgical intervention to secure components or stabilise the structure or bone.
  • 조치
    Johnson & Johnson Medical (JJM) are advising users to remove and return all opened Titanium Low Profile Neuro Screw; Self- Drilling 3mm screws in stock, and to remove and return only affected lots of unopened screws in their original packaging. This action has been closed-out on 25/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Cranial Screw PlusDrive ø 1.6 mm, Self-Drilling, L 3mm (part of the Low Profile Neuro System intended for use in selective trauma of the midface and craniofacial skeleton)Part Number: 400.833Lot Numbers: 9951621, 9955377ARTG Number: 157068