Craniomaxillofacial (CMF) Distraction System (modular family of internal distraction devices to lengthen the mandibular body and ramus) BC Distractor Body, end activated, with Universal Joint for CMF Distractor:Distraction length 15 mm 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Synthes Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00478-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The bc distractor bodies used in the craniomaxillofacial distraction system may reverse post-operatively. reversing occurs when the distractor screw turns in the opposite direction and causes the assembly to lose distraction distance. in the presence of inadequate distraction or reversal of a previously achieved distraction distance, the patient may need to undergo prolonged distraction therapy. in some instances, revision surgery may be needed to replace the device. additionally, there is the potential for those patients with an already compromised mandible anatomy to be at higher risk of partial or complete airway obstruction (critical obstructed airway / choking).
  • 조치
    Synthes Australia is requesting hospitals to immediately isolate all affected units. All units are to be returned to Synthes Australia for replacement or credit. Treating physicians are also being notified of the potential for reversing and the associated risks to patient health. To mitigate the risk of a distractor reversal, patients should be closely followed until a clinical and/or imaging examination confirms the desired outcome. The type of post-operative care and supportive treatment should be determined by the surgical team based on a variety of factors that includes, but is not limited to, the severity of the abnormality, the age of the patient, caregiver environment, ability to protect the airway, maintain nutritional status, and other co-morbidities. For more details, please see http://www.tga.gov.au/safety/alerts-device-craniomaxillofacial-distraction-system-140429.htm . This action has been closed-out on 18/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Craniomaxillofacial (CMF) Distraction System (modular family of internal distraction devices to lengthen the mandibular body and ramus)BC Distractor Body, end activated, with Universal Joint for CMF Distractor:Distraction length 15 mmPart No: 04.315.063Lot No: 6883301Distraction length 30 mm Part No: 04.315.066 Lot No: 7082908Distraction length 40 mm Part No: 04.315.068 Lot No: 7266969; IS10386ARTG No: 183673