da Vinci S Surgical System, IS2000; da Vinci Si Surgical System IS3000 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01067-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-10-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The patient cart is a robotic platform used for da vinci surgery. during normal use the cart can be exposed to small amounts of fluids from the environment (during cleaning for example) without impairing function. improper cleaning techniques involving application of excessive amounts of fluid in a splashing or spraying manner can result in penetration of fluids inside the covers. during surgical procedures there is also a small risk that fluid can be transferred inside the patient cart, for example, biological or i.V. fluids. the system user manual at the time instructed the user to contact intuitive surgical in the cases where components came into contact with liquids. however, there was no specific warning about the consequences of exposure to liquids. this recall action was carried out prior to approval of the recall strategy by the therapeutic goods administration.
  • 조치
    Device Technologies has installed a fluid ingress protection kit on all IS2000 and IS3000 systems.