da Vinci Si and Xi Surgical System - Touchpad Compact Flashcardsda Vinci Si/Xi SSC Touchpad 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00312-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Intuitive surgical has found that compact flash cards used in touchpads on specific da vinci si and xi surgical systems to save user preferences, have the potential to become corrupted when the touchpads experience an abrupt power loss and/or are powered on and off within a short period of time. the corruption of the compact flash occurs when power is cycled. if the compact flash has been corrupted, the da vinci system will display a fault during the start-up of the system. in the worst case scenario, the da vinci surgical system will encounter a non-recoverable fault, preventing use of the system. the error typically presents itself during start-up, so will not likely occur during a procedure (unless a mid-procedure power restart is required).
  • 조치
    Device Technologies is advising users that a representative will be replacing the affected compact flash cards with unaffected products. This may require users to re-enter their user preferences. This action has been closed-out on 27/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    da Vinci Si and Xi Surgical System - Touchpad Compact Flashcardsda Vinci Si/Xi SSC TouchpadPart Number:187505-04 da Vinci Xi PSC TouchpadPart Number: 187716-03 ARTG Number: 97348