da Vinci Si EndoWrist One Vessel Sealer and da Vinci Xi EndoWrist Vessel Sealer da Vinci Si EndoWrist One Vessel Sealer 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00222-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-03-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Intuitive surgical has identified the potential for a certain, small portion of vessel sealers to exhibit interference in the proximal end (near the joint) of the instrument grips. this interference is due to manufacturing variability in some jaws. the interference can lead to contact between the instrument electrodes at the back of the jaws and subsequent shorting of the electrosurgical energy. the shorting behaviour can lead to extended activation times and insufficient energy delivery to thin, vascular tissue bundles, leading to bleeding if transected. this issue has been identified on a small population of vessel sealers only when used on thin, vascular tissue. there has been no impact identified to sealing performance on larger vessels or tissue bundles.
  • 조치
    Device Technologies Australia (DTA) is informing the customers of the potential for insufficient energy delivery on a small population of Vessel Sealers only when used on thin, vascular tissue and providing instructions on alternative methods of sealing and transection. In caseWhere the customer does not wish to use affected Vessel Sealer instruments, DTA is offering a full credit refund for affected stock in its original packaging. This action has been closed-out on 08/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    da Vinci Si EndoWrist One Vessel Sealer and da Vinci Xi EndoWrist Vessel Sealer da Vinci Si EndoWrist One Vessel Sealer Product Code: 410322-05; Order code INS-410322Multiple Lot Numbersda Vinci Xi EndoWrist Vessel Sealer Product Code: 480322-04; Order Code INS-180322Multiple Lot NumbersARTG Number: 132453
  • Manufacturer

Manufacturer