da Vinci Xi 5mm-8mm Universal Cannula Seal and da Vinci Xi 12mm & Stapler Universal Cannula SealUniversal Seals, 5mm-8mm 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00701-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-06-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Device technologies australia (dta) in conjunction with the manufacturer, is issuing this warning notice in relation to the use of da vinci xi 5mm-8mm universal cannula seal and da vinci xi 12mm & stapler universal cannula seal in all surgical procedures.During internal inspections, intuitive surgical has identified particulate in the insufflation stopcocks on the da vinci xi 5mm-8mm universal cannula seal. the particulate generated from the seal material has the potential to be introduced into the cannula lumen when insufflation is connected. there have been no injuries reported related to this issue.
  • 조치
    Users are advised to not use the da Vinci Xi 5mm-8mm Universal Cannula Seal or the da Vinci Xi 12mm & Stapler Universal Cannula Seal for insufflation, or open the stopcock during all types of surgical procedures until this issue is resolved. It is recommended to use an alternative device such as an angio-catheter, Veress needle or laparoscopic trocar for initial access and to maintain insufflation throughout all procedures as an alternative to insufflation through the da Vinci Xi 5mm-8mm Universal Cannula Seal (INS-470361) or the da Vinci Xi 12mm & Stapler Universal Cannula Seal (INS-470380). This action has been closed-out on 20/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    da Vinci Xi 5mm-8mm Universal Cannula Seal and da Vinci Xi 12mm & Stapler Universal Cannula SealUniversal Seals, 5mm-8mmPart Number: INS-740361 Stapler Universal Cannula Seal, 12mm & StaplerPart Number: INS-470380ARTG Number: 181744
  • Manufacturer

Manufacturer