da Vinci Xi EndoWrist Stapler 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01006-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-08-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There has been an increase in the number of complaints regarding the “stapler firing failed” message for the xi stapler. all cases involving the complaints were completed minimally invasively. one complaint of partial fire resulted in a serious injury when a partial fire across an artery resulted in bleeding (1/12900 xi staple procedures = 0.01%), which was resolved through use of pressure, suture, and a clip.The failure message interrupts the stapler firing prior to completion of the full staple line. when this error occurs, the transection of tissue may be incomplete. in all cases formed staples precede the cut line however; there may be unformed staples at the end of the line. the knife blade could be exposed. while this message can occur with gray, white, blue, or green reloads, the greatest concern is with gray and white reloads, where a partial fire could potentially occur while transecting critical vascular structures such as pulmonary, renal or splenic vessels.
  • 조치
    Device Technologies is advising users that In the event of the “Stapler Firing Failed” message, please note the following: 1. While the Stapler remains in the clamped state, it provides compression and tamponades the vasculature. 2. Surgeons should ensure hemostatic and leak control, particularly when around critical vessels with the gray or white reload, before releasing the Stapler from the structure. 3. After unclamping, a blade may be exposed; hence, surgeons should not manipulate tissue with the Stapler after the “Stapler Firing Failed” message occurs. 4. Bedside assistants should take care when removing the fired reload, as sharps hazards, such as potentially exposed knife and unformed staples, may be present. A software correction has been completed as a permanent correction. This action has been closed-out on 03/02/2017.