da Vinci Xi EndoWrist Stapler Release Kit (SRK) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00440-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-04-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, intuitive surgical has received complaints where the srk tool has broken during use. it was determined that misinterpretation of instructions provided on the is4000 srk instruction card may result in damage of the tool. there have been three clinical cases where the endowrist stapler jaws would not open while clamped on tissue, the srk tool broke, and the stapler was excised using an alternative stapling device or other surgical intervention. the risk to health due to the srk tool breakage can range from the minor delay in surgery in order to locate and use a backup srk or instrument release kit (irk) tool, to extracting the clamped tissue from the stapler jaws, to the worst case scenario of the clamped tissue needing to be excised using an alternative stapling device or other surgical intervention. device technologies wish to reiterate the proper use of the srk tool, and to provide additional instructions in the scenario of srk tool damage.
  • 조치
    Device Technologies is advising users in the scenario where the SRK needs to be used, to follow all instructions as provided in the da Vinci Xi EndoWrist Stapler Instruments and Accessories User Manual Addendum and reiterated in the customer letter. If the SRK tool breaks, proceed with the instructions using the Instrument Release Kit (IRK) tool to finish manually unclamping the stapler and contact your Device Technologies representative to arrange for SRK return through the product complaints process.

Device

  • 모델명 / 제조번호(시리얼번호)
    da Vinci Xi EndoWrist Stapler Release Kit (SRK)Product Name: IS4000 EndoWrist Stapler 45 – Stapler Release KitProduct Code: INS-381215Part Number: 381215-02ARTG Number: 228401