Da Vinci Xi Surgical System - Breakage of input disks of Endowrist instruments Reprocessing Instruction: 551708-01 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00560-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-07-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, intuitive surgical inc, has determined that the use of the getinge 88 turbo washer/disinfector in combination with getinge clean mis to reprocess instruments results in a higher chance of input disk breakage, rendering the instrument inoperable. the appendices in the reprocessing manuals incorrectly list the getinge 88 turbo washer/disinfector and the getinge clean mis combination for reprocessing instruments. the getinge combination of equipment and cleaning agents leaves a residual amount of detergent on the input disks resulting in disk breakage during sterilisation.
  • 조치
    Device Technologies Australia (DTA) is requesting the customers to discontinue the use of the Getinge 88 Turbo Washer/Disinfector with Getinge Clean MIS for reprocessing da Vinci Xi instruments. DTA is suggesting alternative reprocessing methods. DTA is also instructing the customers to inspect the input disks of the Endowrist Instruments and identify any broken disks. Device Technologies representative will contact the customer to arrange the returns. This action has been closed-out on 10/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Da Vinci Xi Surgical System - Breakage of input disks of Endowrist instruments Reprocessing Instruction: 551708-01ARTG Number: 97348
  • Manufacturer

Manufacturer