DASH, PDM and Tram models (Multiparameter patient monitor) Manufactured between October 2010 and September 2012 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00258-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-04-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ecg and respiration signal noise caused by dash, pdm and/or tram ecg cable connector movement may result in reduced ecg, respiration and arrhythmia detection performance.
  • 조치
    End users have been provided with work around instructions to help mitigate the issue. GE Healthcare is currently in discussion with TGA as how to best expedite identification of, and prioritised correction of affected units. A follow-up letter will be forwarded once there is a consensus as to the process to be undertaken. This action has been closed-out on 12/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    DASH, PDM and Tram models (Multiparameter patient monitor) Manufactured between October 2010 and September 2012Dash 3000/4000/5000Product codes: SHQ and SD0Patient Data Module (PDM)Product code: SA3Tram 451, 451N, 451M, 851, 851N, 451N5, 851N5Product codes: RT1, RT4, RT5, RT7, RT8, RTW, SCE and SBBARTG Number: 92726
  • Manufacturer

Manufacturer