Defibrillator/pacer SLIM corpuls3 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00979-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-09-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Due to too little distance between a circuit board and the housing, there is the possibility that a component on that circuit board could be damaged mechanically. this impairs the radio connection between the modules. the user can recognise this, if a respective error message (ie, “no connection to defibrillator unit” or “no connection to p-box” ) is displayed. impairment of the radio connection between the modules could result in the inability to: - perform full patient monitoring; and - correctly adjust and operate a therapy function of the device. vibrations have been found to increase the likelihood of this malfunction. due to this issue, diagnosis and therapy are delayed, because the measurement and therapy functions cannot perform without problems.
  • 조치
    Device Technologies Australia (DTA) is advising the customer to connect the module mechanically to form a compact device if the error message is displayed. A permanent correction of the error is only possible by replacing the affected radio module. DTA is arranging replacement of the radio module. This action has been closed-out on 1/06/2016.