Dexcom G4 PLATINUM and Dexcom G5 Mobile receiver (Continuous Glucose Monitoring Device). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00224-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Dexcom noticed an increase in complaints related to audible alarms and alerts associated with dexcom g4 platinum and dexcom g5 receivers.As such, all dexcom customers are being notified that that may not receive an intended audible alarm or alert. as a result, if they rely on hearing the alarm or alert, they may not detect a severe hypoglycaemic (low glucose) or hyperglycaemic (high glucose) event. this includes the fixed low alarm at 3.1 mmol/l, which alerts the user with vibration first, followed by beeps 5 minutes later if not confirmed. the vibration alert should continue to work; however, the audible beeps may not function. the vibration function for the alerts and alarm cannot be disabled by the user.
  • 조치
    Emergo is providing customers with instructions on how to test the audible alarms to ensure they are functioning properly. If the alarms and alerts are not functioning, users are advised to contact the local distributor, Australasian Medical and Scientific Ltd (AMSL), for a replacement device. Dexcom is working on implementation of an improved speaker for the receivers. For more information please see https://www.tga.gov.au/alert/dexcom-continuous-glucose-monitor-receivers . This action has been closed-out on 10/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Dexcom G4 PLATINUM and Dexcom G5 Mobile receiver (Continuous Glucose Monitoring Device). Part Number: MT20649-2, MT22719-2ARTG Number: 169241
  • Manufacturer