Dimension and Dimension Vista Clinical Chemistry Systems. An in vitro diagnostic medical device (IVD) Assays affected: Triglycerides, Enzymatic Creatinine, Automated HDL Cholesterol, Uric Acid and HDL Cholesterol 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00350-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has become aware of n- acetylcysteine (nac) and metamizole (dipyrone) interference with trinder and trinder-like reaction assays. the trinder reaction is a reaction where hydrogen peroxide is formed and subsequently reacts with a phenol derivative and aminoantipyrine in the presence of peroxidase to form a coloured quinone product. siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving n- acetylcysteine (nac) or metamizole. nac is the accepted antidote for acetaminophen toxicity and is justified in patients at significant risk for hepatotoxicity. the potential exists for reporting falsely depressed results for the assays affected. baseline values before administration of the nac or metamizole therapy would not be affected. it is extremely unlikely these assays would be requested during assessment of acetaminophen overdose and nac treatment.
  • 조치
    Siemens is advising the IFUs for the relevant assays will be updated to indicate that "Venipuncture should occur prior to Metamizole/N-Acetyl Cysteine (NAC) administration due to the potential for falsely depressed results." A review of previously generate results is at the discretion of the Laboratory Director.