Dimension Clinical Chemistry System and Dimension Vista System. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00839-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-07-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has become aware of sulfasalazine and sulfapyridine drug interference in certain assays which use nadh and/or nadph to generate reduction oxidation reactions which produce colorimetric signals.Siemens has confirmed that falsely depressed or falsely elevated results may occur on samples drawn from patients taking sulfasalazine and sulfapyridine. sulfasalazine is the accepted treatment for inflammatory bowel disease, ulcerative colitis, crohn’s disease, rheumatoid arthritis, inflammatory arthritis, and uveitis. sulfapyridine is used occasionally for dermatitis herpetiformis and related skin disorders when alternative treatment is unsuitable.Due to this drug interference, siemens will update the limitations of the procedure section of the instructions for use (ifu) for specific dimension and dimension vista assays.
  • 조치
    Siemens is requesting that Customers: 1. Review this letter with their Medical Director; 2. Venipuncture should occur before drug administration of Sulfasalazine or Sulfapyridine as indicated under Reason/Issue. Baseline assay values before administration of Sulfasalazine or Sulfapyridine therapy would not be affected; 3. If any complaints of illness or adverse events are received associated with the specified products within the Customer Letter immediately contact your local Siemens Customer Care Centre or your local Siemens technical support representative; 4. Complete and return the Response Form by Friday 4th August 2017 as confirmation that you received and understood the bulletin; and 5. Retain the letter with your laboratory records, and forward the letter to those who may have received this product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Dimension Clinical Chemistry System and Dimension Vista System. An in vitro diagnostic medical device (IVD)Assays: Ammonia, Alanine Aminotransferase, Aspartate Aminotransferase, Glucose, Creatine Kinase MB and Thyroxine.Multiple Test Codes and Catalogue NumbersAll LotsARTG Numbers: 181689, 181688, 182776, 180109
  • Manufacturer

Manufacturer