Dimension Enzyme II Calibrator (used with Dimension clinical chemistry system). An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00661-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-05-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has received complaints of positive shift of patient and qc results when the alti assay is calibrated with the affected dimension enzyme ii calibrator lots. internal investigation has confirmed that patients, quality control, and proficiency testing material all demonstrate a similar shift of up to 10% in comparison to results obtained using earlier calibrator lots. there is no risk to patients however there may be a delay in results.
  • 조치
    Siemens has reassigned the bottle values of all Enzyme II calibrator lots currently in date. Laboratories may expect to observe a downward shift of approximately 8% in QC, patient samples and proficiency testing materials when reassigned bottle values are used for calibration. Siemens recommends recalibration of existing Alanine Aminotransferase (ALTI) flexes and future calibrations using the reassigned calibrator bottle values for lots provided. A review of previously generated results due to this issue is at the discretion of the laboratory.

Device

  • 모델명 / 제조번호(시리얼번호)
    Dimension Enzyme II Calibrator (used with Dimension clinical chemistry system). An in vitro diagnostic medical device (IVD)Product Code: ENZ II CALCatalogue Number: DC143Siemens Material Number (SMN): 10476170Multiple Lot Numbers and Expiry Dates affectedARTG Number: 181688
  • Manufacturer

Manufacturer