Dimension Vista 500 and Dimension Vista 1500 using software versions 3.5.1 and 3.6 - Clinical chemistry analyser. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00649-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-06-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare has confirmed that, under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a qc or calibrator product from a vista vial. if two samples are dispensed into the same aliquot well, the first sample will always be qc or calibrator from a vista vial. patient sample will never be dispensed into an aliquot well that contains another patient sample.
  • 조치
    In the event of an instrument reset, end users are requested to check if there are any QC or calibration tests pending. If there are QC or calibrations pending a software restart is necessary before resuming patient processing. If there is no QC or calibration pending reset or troubleshoot the instrument as per the Operators Guide. A software update will be provided as a permanent fix.

Device

  • 모델명 / 제조번호(시리얼번호)
    Dimension Vista 500 and Dimension Vista 1500 using software versions 3.5.1 and 3.6 - Clinical chemistry analyser. An in vitro diagnostic medical device (IVD)Siemens Material Numbers: 10488224 and 10444802ARTG Number: 174699