Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR750w GEM products (Full body magnetic resonance imaging systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00385-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Gradient coil electromechanical connections may loosen causing an increase in resistance. the increased resistance can lead to increasing heat during scanning which, in turn, can cause material in the rear endbell enclosure to overheat, generate smoke in the scan room, and cause some localised melting of the endbell. this overheating has occurred only while units were being serviced by ge service engineers during repeated evaluation scanning. however, the issue could also occur in clinical mode.
  • 조치
    While in clinical mode, If the operator is notified via user interface message that there are multiple X, Y, or Z gradient distortion faults they should monitor for smoke odours or evidence of overheating. If they detect either a smoke odour or evidence of overheating they should stop scanning immediately, remove the patient and call their GE Healthcare Service Representative. Scanning should not resume until the GE Healthcare service representative has confirmed the MRI scanner is functioning normally. GE Healthcare will correct all affected products.