Discovery MR450, MR750 and Optima MR450w MRI System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00885-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-08-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has recently become aware of a potential safety issue due to exposed yellow foam on the end bells of discovery mr450, mr750 and optima mr450w product. the yellow foam strip used between the end bell covers and patient bore of certain ge mr products may extend past the surface of the covers. the foam itself is not hazardous. however, any fluid or body substance absorbed by the extended foam may come into contact with patients and act as a fomite potentially transferring infectious agents from one patient to another.
  • 조치
    GE Healthcare is advising users that the system may continue to be used as normal even with the yellow foam protruding from the bore. Patient padding should be used as described in the Operating manuals and cleaning practices described in the Operating manuals should be followed. Patient padding will minimise contact with the foam and users are advised not to attempt to remove the foam. GE Healthcare will be installing a permanent cover over the foam in October 2013.

Device

  • 모델명 / 제조번호(시리얼번호)
    Discovery MR450, MR750 and Optima MR450w MRI SystemARTG Number: 108415