Discovery MR450w, Discovery MR750, Optima MR450w, Optima MR450w equipped with the GEM option, Discovery MR750w and Discovery MR750w equipped with the GEM option products (Full body magnetic resonance imaging (MRI) systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00464-2
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has recently become aware of a potential safety issue with the patient table cradle emergency release handle and cradle release block mechanism of certain ge mri products. patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block mechanism malfunctions or becomes misaligned. this could make difficult the safe removal of the patient from the system.
  • 조치
    GE Healthcare will correct all affected products. Before continued use prior to repair, GE is advising that the operator should assess patient positioning and ease of egress and cease use if cradle movement for egress is impeded. If the operator encounters difficulty in moving the cradle out because the cradle release handle does not release properly or if the cradle movement is impeded, the facility’s GE Healthcare Service Representative should be contacted so that the cradle can be inspected and necessary repairs can be made as soon as possible. This action has been closed-out on 02/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Discovery MR450w, Discovery MR750, Optima MR450w, Optima MR450w equipped with the GEM option, Discovery MR750w and Discovery MR750w equipped with the GEM option products (Full body magnetic resonance imaging (MRI) systems)ARTG Number: 108415
  • 의료기기 분류등급
  • Manufacturer

Manufacturer