Drager Infinity Acute Care System (IACS) Monitoring Solution with software versions VG2.2 to VG6.0 and the Infinity M540 Stand-Alone Patient Monitor with software versions from VG2.2 to VG6.0 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00154-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-03-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Drager has determined that when the infinity m540 patient monitor is placed in standby mode and the user simultaneously connects the infinity mcable- dual hemo or infinity mpod quad-hemo to the infinity m540 patient monitor and then proceeds to zero the invasive pressure channel, in rare cases the invasive pressure visual and audible alarms are not activated. although the invasive pressure value displayed on the infinity m540 patient monitor and on the infinity medical cockpit is accurate, the invasive pressure visual and audible alarms will stay disabled until the patient is discharged or the hemodynamic pod is disconnected and reconnected.This issue is not seen when the user puts the m540 into stand-by mode, and then connects the hemodynamic pod in a sequential manner. the actual ibp measurements, and the patient values are correctly displayed on both the m540 and cockpit at all times.
  • 조치
    Drager is advising users to follow the below identified steps whenever the Hemodynamic Pod is plugged into the Infinity M540 Patient Monitor and the unit is placed in Standby. When users are ready to monitor the patient: 1. Exit Standby Mode 2. Unplug, pause for a minimum of 3 seconds and re-plug the Hemodynamic Pod 3. Continue per the normal procedure. Drager will be performing a software update as a permanent correction.

Device

  • 모델명 / 제조번호(시리얼번호)
    Drager Infinity Acute Care System (IACS) Monitoring Solution with software versions VG2.2 to VG6.0 and the Infinity M540 Stand-Alone Patient Monitor with software versions from VG2.2 to VG6.0All models affectedARTG Numbers: 168054 and 98600
  • 의료기기 분류등급
  • Manufacturer

Manufacturer