Drager V Series Ventilator with the Optional PS500 battery Installed 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01150-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-11-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall is due to the unexpected shortening of the optional ps500 battery on an iacs workstation critical care unit. the batteries were discharged early, even though the device was indicating an adequate battery state of charge. the devices generate a "battery low" alarm, followed by "battery flat". in exceptional cases, the "battery flat" alarm might be displayed too late, or not at all. after complete discharge of the batteries, the device correctly alerts to failure of the power supply by the auxiliary alarm (piezo, power failure alarm) in all cases. it has been discovered that the current design of the charging algorithm for the new ps500 batteries is not able to appropriately prevent premature battery ageing in all application scenarios.The unexpected loss of automatic ventilation functionality due to depleted battery requires immediate user intervention to prevent serious patient harm, including death. no patient injuries have been reported so far.
  • 조치
    The sponsor Draeger is advising users not to use devices with the affected PS500 battery unit for patient transport until the firmware has been updated.

Device

  • 모델명 / 제조번호(시리얼번호)
    Drager V Series Ventilator with the Optional PS500 battery InstalledAffected models:Infinity V500Babylog VN500Evita V300All serial numbers affectedARTG Numbers: 158739, 170141 and 92177
  • Manufacturer

Manufacturer