Drager V Series Ventilator with the Optional PS500 External Power Supply Unit installed 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00176-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-02-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Dräger became aware that the battery capacity of the optional ps500 power supply unit was unexpectedly reduced. this resulted in batteries prematurely discharging, even though an adequate battery charge status was displayed. the "battery discharged" alarm was not in all cases triggered five minutes prior to battery discharge, as specified. when the batteries were completely discharged, the devices triggered the existing power failure alarm correctly. the reported issue may cause unexpected loss of automatic ventilation functionality.Further to the recall for product correction in feb 2016 (tga ref.: rc-2016-rn-00154-1), dräger has now developed device modifications to rectify the issue.
  • 조치
    The device modifications include the following:  - Installation of a new software 2.50 with improved battery charge management  - Improved charge management for external batteries  - Option of performing a battery check to determine the battery's current runtime  - Option of displaying the remaining battery capacity in minutes  - Installation of an additional internal backup battery  - Replacement of the power supply unit to support both the external battery and an additional internal battery  - Replacement of the existing external batteries. The installation of a backup battery will ensure that the internal battery will continue to supply power should the external battery fail. Users are advised to continue to observe the information in the Feb 2016 letter until the devices have been updated with the improved system.