Drager V Series Ventilator with the Optional PS500 Installed 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00154-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-02-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The battery capacity of the optional ps500 power supply unit for the evita infinity v500, evita infinity v300 and the babylog vn500 rapidly and unexpectedly reduces despite using the most recent power supply firmware (fw1 .50) which was released in november 2015 .This resulted in batteries prematurely discharging, even though an adequate battery charge status was displayed. the "battery discharged" alarm was not in all cases triggered five minutes prior to battery discharge, as specified. when the batteries were completely discharged, the devices triggered the existing power failure alarm correctly.The reported issue may cause unexpected loss of automatic ventilation functionality which could lead to serious injuries to the patient. so far no injuries have been reported.
  • 조치
    The sponsor is cancelling the distribution of the updated power supply firmware (FW1.50) and rolling back devices on which this firmware version is already installed to FW1.49. As a preventive measure, Dräger will replace the batteries of all PS500 customers every 6 months until a permanent technical solution is available. Dräger is also advising users NOT to use the device for patient transport unless absolutely necessary and NOT to rely on the battery status indicator.