Drager Ventstar Paediatric Circuit, used in conjunction with the Dräger Oxylog 3000plus Transport Ventilator 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00721-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-05-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During the course of routine internal testing drager detected isolated leakages at the check valve (one way valve of the inspiration branch) of the above mentioned disposable paediatric ventilation circuit. leakage at the check valve can result in a patient's exhaled gas entering into the breathing circuit, which could lead to the rebreathing of the exhaled gas with subsequent reduced oxygen concentration for the patient. this leakage is not detected during the ventilator operational readiness check. to date, drager has not received any complaints associated with this issue.Further detailed investigations have shown that the check valve functions properly with no leak observed at positive end-expiratory pressures above 5 mbar i cmh20. the problem only occurs at peep values below 5 mbar i cmh20.
  • 조치
    Drager urgently recommends that users immediately inspect any stock and dispose of any product (part number 5704964) with a date of manufacture up to and including 03-2016. A Free of charge replacement product can be obtained by following the instructions in the customer letter.