e.cam and all Symbia systems with foresight detectors (Nuclear medicine gamma camera systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00419-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has determined that e.Cam or symbia systems with foresight detectors performing gated or dynamic acquisitions may lose some detector time-information. the resulting risk is that the framing of the image data and the calculated ejection fraction values may be lower. although misdiagnosis could be possible if ejection fraction values are used as a sole input to diagnosis. in the lifetime of the product, siemens have received three complaints, none of which led to misdiagnosis or adverse events.
  • 조치
    A software patch (Ml16-001) and corresponding operator manual addendum will be implemented. Once the software patch is installed on the affected system, the improvements will further reduce the likelihood of the issue occurring. In the interim, users are advised they may continue to use your system since the probability of occurrence is remote. To reduce the likelihood of this issue, they are advised to perform Daily Shutdown and Start-up as described in the operator manual. This action has been closed-out on 26/05/2017.