Elecsys Thyroid Stimulating Hormone (TSH) and Elecsys Parathyroid Hormone (PTH) (1-84) when used on the cobas e 602 module. An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00256-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-03-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An internal error resulted in assignment of the same lot number for elecsys tsh reagents & elecsys pth (1-84) reagents. the number of tests and qc recovery of elecsys tsh lot 185522 may be affected if the elecsys pth (1-84) lot 185522 was previously used on the same cobas e 602 module. the following may occur: a reduced number of determinations for elecsys tsh assay (100 instead of 200 tests/kit); low recovery of level 2 from precicontrol universal (all lots) < -3 sd; and high recovery of precicontrol thyro sensitive & precicontrol tsh above +3 sd.This issue could cause tsh results to appear above or below the normal ranges. it cannot be excluded that a delayed or incorrect diagnosis would not occur in certain high-risk patient populations where only tsh results are used e.G. in newborn screening or pre-interventional diagnostic scenarios, leading to inadequate or inappropriate treatments. no customer complaints regarding this issue have been received by roche diagnostics to date.
  • 조치
    Roche is advising users to discontinue use of Elecsys TSH reagent lot 185522 on cobas e 602 modules that have previously used Elecsys PTH (1-84) reagent lot 185522. Replacement Elecsys TSH reagents will be provided. It is recommended that users discuss with their medical director whether a look-back is required for previously affected patient results. If hyperthyroidism is excluded based on TSH results only, patients whose results were generated from lot 185522 Elecsys TSH and in the range of 0.26 – 0.36 µIU/mL should be re-tested, especially if the patient’s clinical symptoms point to hyperthyroidism.

Device

  • 모델명 / 제조번호(시리얼번호)
    Elecsys Thyroid Stimulating Hormone (TSH) and Elecsys Parathyroid Hormone (PTH) (1-84) when used on the cobas e 602 module. An in vitro diagnostic medical device (IVD). Elecsys TSH Material Number: 11731459122Lot Number: 185522Elecsys PTH (1-84)Material Number: 05608546190Lot Number: 185522ARTG Number: 174907
  • Manufacturer

Manufacturer