Elekta Monaco rel. 3.10.00 through 3.20.00 & Focal rel. 4.64.00 through 4.70.00 (Radiation Therapy Treatment Planning System)Elekta Monaco rel. 3.10.00 through 3.20.00 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00670-3
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-07-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    If the dicom coordinates option is turned on in the reports pull down menu the following issues are seen on the resulting print out:issue 1: the coordinates will be in dicom space, thus not matching the setup reference dialogue.Issue 2: the shift coordinates are reported incorrectly. the z coordinate should become dicom negative y and the y coordinate should become dicom z.Incorrect shift coordinates are generated based on issue 2 above. these shifts can lead to incorrect patient positioning prior to treatment position verification.
  • 조치
    Customers to implement the work around instructions provided by the sponsor until the problem is resolved with the future release of a software patch. This action has been closed-out on 12/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Elekta Monaco rel. 3.10.00 through 3.20.00 & Focal rel. 4.64.00 through 4.70.00 (Radiation Therapy Treatment Planning System)Elekta Monaco rel. 3.10.00 through 3.20.00ARTG Number: 186856Focal rel. 4.64.00 through 4.70.00ARTG Number: 186853
  • Manufacturer

Manufacturer