Elements Obturation Unit with Buchanan Heat Pluggers 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Kerr Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00057-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-02-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Sybron endo received a recent increase in complaints relating to overheating buchanan heat pluggers only while used with the elements obturation unit (eou). when used in the eou, the affected buchanan heat pluggers exceed sybron endo internal specification and operating temperature specifications provided in the instructions for use (ifu). the heat pluggers tips are intended to heat to 200°c ± 50°c while in down-pack and hot pulp mode, 100°c ± 25°c while in back fill mode and 600°c ± 100°c while in cautery mode. internal testing demonstrated that some tips may exceed 700 °c upon activation, while in any operating mode (down pack, back fill, hot pulp test and cautery mode). when this occurs, the screen will indicate niti and the tip will glow red shortly after activation.
  • 조치
    Kerr Australia is requesting users: 1. Inspect your device immediately and quarantine affected stock as outlined in the item numbers above to prevent further use; and 2. Examine current inventory for affected product and check the heat plugger to see if the NiTi icon appears and/or tip turns glowing red. (see instructions below); and 3. If you observe the NiTi icon and/or glowing red heat plugger tip, do not use heat plugger and return to Kerr. Please call customer care; and 4. Always follow the Instructions for Use when using the product

Device

  • 모델명 / 제조번호(시리얼번호)
    Elements Obturation Unit with Buchanan Heat Pluggers Part Numbers: 973-0322, 952-0004, 952-0003, 952-0005, 952-0006, 952-0007, 952-0031, 973-0219(hot pulp test tip)ARTG Number: 236654 (Kerr Australia - Endodontic obsturation system)