Endoscopic Curved Needle Driver, LapSac Introducer and LapSac Introducer Set 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 William A Cook Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00584-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-05-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Cook medical have identified that the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilisation of specific endoscopic curved needle driver, lapsac introducer and lapsac introducer set products. preliminary investigation indicates that validation data related to the reprocessing of these devices do not meet the current guidance. to date, the manufacturer has received no reports of adverse reactions related to inadequate cleaning, disinfection, or sterilisation associated with these devices. potential adverse events that may occur if the products are not adequately reprocessed include localised surgical site infection, to deeper organ space infection, as well as chemical residual exposure.
  • 조치
    Cook Medical is advising users to examine their inventory immediately to identify and quarantine any affected products. Products are to be discarded or returned to Cook Medical.

Device

  • 모델명 / 제조번호(시리얼번호)
    Endoscopic Curved Needle Driver, LapSac Introducer and LapSac Introducer SetEndoscopic Curved Needle DriverCatalogue Numbers: 096511, J-ENH-033100, J-ENH-053100, J-ENH-053120 and J-ENH-053130ARTG Number: 212233LapSac IntroducerCatalogue Number: J-LSI-102500ARTG Number: 276893LapSac Introducer SetCatalogue Number: 054300Previous ARTG Number: 276893All Lots
  • Manufacturer

Manufacturer