enGen Laboratory Automation Systems configured with Data Innovation, LLC Instrument Manager (IM) v8.06 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Ortho-Clinical Diagnostics 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00414-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The engen laboratory automation system is an automated track which integrates clinical analysers and the customer’s lis (laboratory information system). it centrally controls pre-analytical and post-analytical processing as well as sample transport to multiple independent analysers and devices. ortho clinical diagnostics (ocd) internal testing has determined that when a sample requires follow-up action, such as dilution or retesting due to a mechanical issue, it is possible for that sample to be routed to the sample storage drawer (storage) prior to the completion of the necessary pending action. a delay in patient test result(s) reporting could occur.
  • 조치
    Laboratories are advised to manually rerun samples that have been prematurely stored. A software modification is currently in development and will be provided once complete. This action has been closed-out on 11/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    enGen Laboratory Automation Systems configured with Data Innovation, LLC Instrument Manager (IM) v8.06Product Code: 6802883Lot Numbers: enGen Select Versions 3.2.2 and below & enGen Custom versions 3.2.2 and below