enGen Laboratory Automation Systems (enGen) with TCAutomation Software v2.6, v3.2 and v3.5 Configured with Bypass Modules for ADVIA Centaur, Abbott Architect, VITROS and/or Tosoh AIA Systems (an in vitro diagnostic medical device (IVD)) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Ortho-Clinical Diagnostics 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00485-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ortho clinical diagnostics (ocd) determined that, due to a software timing issue, a sample tube carrier being released in the bypass module may not be detected, potentially resulting in the next sample to be metered not advancing for processing. no error message will be posted if this occurs. the bypass module will stop processing samples indefinitely until an operator detects the issue and intervenes.
  • 조치
    Ortho-Clinical Diagnostics will provide a resolution for this issue in a future software version. In the interim, end users are requested to routinely monitor the Bypass module for stoppage due to the software timing issue. This action has been close-out on 11/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    enGen Laboratory Automation Systems (enGen) with TCAutomation Software v2.6, v3.2 and v3.5 Configured with Bypass Modules for ADVIA Centaur, Abbott Architect, VITROS and/or Tosoh AIA Systems (an in vitro diagnostic medical device (IVD))Multiple affected bypass modules (product codes)
  • Manufacturer

Manufacturer