Engstrom Ventilator & Aespire View, Aisys and Avance Anaesthesia MachinesManufactured from 23 April 2013 to 22 July 2013 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00953-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-09-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has recently become aware of a potential safety issue involving unresponsive buttons on the display of the engstrom ventilator and the aespire view, aisys and avance anaesthesia machines. as a result of a manufacturing issue, the buttons on the left, right and bottom keypads may not always detect button presses. this may result in the inability to access certain menu functions which could possibly lead to delay in treatment.
  • 조치
    The user may notice that a button press may not actuate the desired function. Usually by pressing the key again, the device will respond as expected. If a button has no response, the menu function may be accessible through the use of the control wheel. If none of the above steps can be performed, discontinue use and contact a GE Healthcare Service Representative. GE Healthcare will correct all affected systems.

Device

  • 모델명 / 제조번호(시리얼번호)
    Engstrom Ventilator & Aespire View, Aisys and Avance Anaesthesia MachinesManufactured from 23 April 2013 to 22 July 2013Multiple serial numbers affectedARTG Number: 93955
  • Manufacturer

Manufacturer