EPIQ 7 Ultrasound system version 1.0 (Intended for diagnostic ultrasound imagingand fluid flow analysis) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01011-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-09-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A problem has been detected with philips model epiq 7 ultrasound system v1.0 when using philips qlab a2dq software feature. it is possible to have erroneous ejection fraction (ef) measurements appear in the biplane analysis report page under certain conditions while using a2dq. the biplane analysis report page could display ap2 or ap4 edv/esv/ef measurements from a prior patient’s exam. erroneous measurement information in the biplane analysis report page could mistakenly influence patient care. this issue only occurs on software version 1.0 in a2dq during a live (frozen) image and only when a single biplane view is included in the exam. when both biplane views are present (ap2 and ap4) this issue does not occur.
  • 조치
    Philips corrected this issue in Version 1.01 of the EPIQ 7 Ultrasound System. A Philips Field Service Engineer will be contacting customers to schedule the V1.01 upgrade to affected systems. Philips is providing temporary workaround instructions to prevent the failure mode from occurring until the systems are upgraded.

Device

  • 모델명 / 제조번호(시리얼번호)
    EPIQ 7 Ultrasound system version 1.0 (Intended for diagnostic ultrasound imagingand fluid flow analysis)ARTG Number: 93851
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA