Euphora Semi-Compliant Rapid Exchange Balloon Dilatation Catheter 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00862-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-07-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medtronic has received reports of difficulties in removing the stylette of the euphora semi-compliant rapid exchange (rx) balloon dilatation catheter. in the event that stylette removal difficulties are encountered, there is the potential for damage to the balloon or catheter due to excessive force being applied, and this may lead to balloon inflation or deflation challenges. potential risks associated with balloon inflation and deflation difficulties include prolonged procedure time and the need for additional intervention. in addition, inflation or deflation difficulties or catheter damage during procedural use could result in vessel occlusion or injury. medtronic has received one report of patient injury related to this matter.
  • 조치
    Medtronic is requesting customers: 1. Identify and quarantine all unused affected product as listed in your inventory; 2. Return all unused, affected product listed in your inventory to Medtronic. Contact Customer Service on 1800 668 670 to initiate a product return and credit. Your Medtronic Representative can assist you in the return of affected product as necessary; and 3. Complete the supplied Customer Acknowledgement Form and fax it to Medtronic on 1800 929 645 to the attention of ANZ Regulatory Affairs or scan and email to rs.anzregaffairs@medtronic.com with TGA Ref # “RC-2017-RN-XXXXX-X” in the subject line.

Device

  • 모델명 / 제조번호(시리얼번호)
    Euphora Semi-Compliant Rapid Exchange Balloon Dilatation CatheterLot Number range: 212553130 to 213283557Expiration Date range: 4 Jan 2019 to 9 May 2019Multiple Model NumbersARTG Number: 238388
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA