ExacTrac versions 6.0.1 and 6.0.2 only (Image Guided RadiationTherapy Patient Positioning System) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Brainlab Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00218-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-03-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    After performing a snap verification, that is a monoscopic intra-fraction x-ray monitoring using only one x-ray tube, in combination with implanted markers, the exactrac 6.0 x-ray monitoring tab might display an incorrect hint icon and result text. exactrac might incorrectly show the hint icon in green colour with the result text “no deviation detected”, despite the simultaneously displayed unaffected x-ray image showing the actual positions of the implanted markers potentially deviating from the user defined tolerance areas .
  • 조치
    Brainlab is asking customers to always verify the result of a Snap verification within the simultaneously displayed x-ray images generally required and never exclusively rely on the hint icon colour and the result text only. Brainlab will provide a software update to affected customers around end of August 2013 to correct the issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    ExacTrac versions 6.0.1 and 6.0.2 only (Image Guided RadiationTherapy Patient Positioning System)ARTG: 121226