Exeter Small and Large Tapered Pin Reamer (used as part of the Exeter Procedure Pack) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00972-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-10-01
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker has received a customer complaint which reported that during a total hip replacement the surgeon was using the exeter small tapered pin reamer when the device fractured approximately 3 inches from the tip. this may result in the tip of the reamer detaching from the instrument. should this occur and the fragment remain in situ, there are additional potential hazards including excessive metal ions and displacement from mri.
  • 조치
    Stryker is advising users to quarantine any affected stock for replacement.

Device

  • 모델명 / 제조번호(시리얼번호)
    Exeter Small and Large Tapered Pin Reamer (used as part of the Exeter Procedure Pack)Catalogue Numbers: 0932-0-000 (small) and 0932-2-000 (large)Lot Numbers (small): GW281208, GW284192, GW288907, GW282586, GW282310, GW298838, GX303720, GX302994 and GX301732 Lot Numbers (large): GW288908, GW282454 and GW288485ARTG Number: 140892
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA